PERFORMANCE |
|
|
Compatible sensors |
Strain gauge transducer |
|
Signal input terminals |
Round connector (NDIS7P)/terminal bank (connect only one at a time) |
|
Operating temp range |
0°C to 40°C |
|
Storage temp range |
-20°C to 60°C |
|
Excitation |
2.5 VDC ±5% (30mA maximum current) |
|
Signal input range |
±5 mV/V |
|
Weight |
320 g |
Including batteries |
D/A Output |
± 2V |
|
A/D Conversion |
1000 times/second, 24-bit |
|
Digital filter |
Moving average (select from OFF, 16, 32, 64, 128, 256, 512, 1024, 2048) |
|
TEDS function |
IEEE1451.4 class 2 mix mode interface |
|
Hold functions |
Sample hold, peak hold, bottom hold, zone definition hold (peak, bottom) |
|
Power supply |
4 alkaline or NiMH AA batteries USB bus power (built-in Micro-USB B connector) |
|
Operating humidity range |
85% RH or less (without condensation) |
|
Applicable standards |
CE marking, VCCI (Class A), FCC (Class A) |
|
External dimensions (WxHxD) |
Approximately 85mm x 140mm x 35mm (without protrusions) |
|
EQUIVALENT INPUT/TEDS |
|
|
Calibration range |
0.3 mV/V - 0.5mV/V |
|
Calibration precision |
Within 0.1% F.S. (when using a 1m standard HITEC Ø8, 6-core shielded cable with 350Ω impedance, when 5mV/V) |
|
PRECISION |
|
|
Non-linearity |
Within 0.01% F.S. + digit (when 5mV/V) |
|
Zero drift |
Within 0.5μV/°C (input conversion value) |
|
Gain drift |
0.005% / °C or less |
|
D/A OUTPUT |
|
|
Output connector |
BNC |
|
Output voltage |
±2.0V |
|
Resolution |
70.16μV typ |
|
Non-linearity |
0.02% F.S. or less |
|
Zero drift |
0.1mV/°C or less |
|
Gain drift |
0.003%/°C or less |
|
DISPLAY |
|
|
Display |
2.4" color TFT LCD |
|
Display modes |
Setting screens, indicator value digital display, graph display, recorded data list display, static strain display |
|
Languages |
Japanese/English |
|
INDICATOR VALUE |
|
|
Display range |
-99999 to 99999 |
|
Decimal point |
Display position selectable |
|
DISPLAYED ITEMS |
|
|
Calibration settings |
Zero calibration/span calibration (TEDS calibration, actual load calibration, equivalent input calibration) |
|
Function settings |
High limit, low limit, comparison mode, hysteresis, nearly zero, moving average, motion detect, zero tracking, digital zero, digital zero offset, zone definition |
Hold mode, control lock, minimum grid, digital zero limit, clear digital zero, select data output, D/A converter |
Sensor value memory |
Six types of calibration values for each connected sensor |
|
DATA RECORDING |
|
|
Indicator value |
300 maximum. Recorded contents: ID number, date and time, recording mode, sensor value memory number, indicator value |
|
Graph recording |
8 maximum. Recorded contents: ID number, date and time, sensor value memory number, trigger mode, grapgh waveform |
|